- Contact Person : Mr. yan juezhi
- Company Name : Wuxi Guanya Refrigeration Tech Co., Ltd.
- Tel : 86-510-88321168
- Fax : 86-510-88587187
- Address : Jiangsu,wuxi city,maqiao industril zone,hongsheng street ,new strict of wuxi
- Country/Region : China
- Zip : 214112
closed refrigerated and heating thermostat -25 ~ 180 degree
Description:
Heating and cooling circulator is mainly applied to the glass reactor and precisely control the temperature and process of the reactor.It can keep the the temperature constant at the setpoint temperature.
1.The whole system is a closed liquid circulation system with the expansion of container, expansion of the container and the liquid circulation is adiabatic, and did not participate in liquid circulation, only mechanical connection, regardless of the temperature of liquid circulation is a high temperature or low temperature expansion of the medium container of low at 60 degrees.
2.Throughout the liquid circulation system is sealed, low temperature did not absorb water vapor, high temperature did not produce the oil miss
3.The entire circulatory system is not used in machinery and electronic valves.
4.Equipped with integrated heating and cooling the container, the rate of warming and cooling fast, heat-conducting oil demand is relatively small.
5.Continuous heating and cooling control
6.The cycle is closed, high temperature oil mist was not volatile, will not be conducting oil oxidation and browning; low temperature will not absorb water vapor in the air; extended the life span of Heat Transfer Oil
7.With self-diagnosis function, freezer overload protection, high pressure switches, overload relays, thermal protection device and other security functions, and fully guarantee the use of safety
8.Taikang compressor used compressors of domestic law or the United States and Germany Compressor gifted musicians.
9.correction loop with a temperature probe PT100 function.
10.HCFE and CFC-free refrigerants.
11.High-lift design to meet the long-distance transmission medium heat.
Heating and cooling circulator | |
Model | SST-15 |
Working temp range | —25°C~180°C |
Controller | PID |
Temperature accuracy | ±0.5°C |
Heating power | 1500W |
Cooling power | 1000W AT 180°C 800W AT 50°C 500W AT 0°C 415W AT —10°C 280W AT —20°C |
Pump L/min bar | Max15L/min 0.8BAR |
Input, display | LED display, touch-button input |
Security protection | Self-diagnostic function; freezer overload; high-pressure pressure switch, overload relays, thermal protection device and other security features |
Refrigerants | R-404A |
Outlet | DN15 |
Return | DN15 |
Dimensions | 260×450×580 mm |
weight | 45kg |
Power | AC 220V 50HZ 2000W(max) |
Case material | SUS 304 |
closed refrigerated and heating thermostat -25 ~ 180 degree